Many of our customers face these challenges - Yokogawa Chlorine Analyzers resolve them.
The FC800D non-reagent free available chlorine analyzer adopts the polarographic method, and a unique rotating electrode provides a continuous online measurement of free available chlorine concentration.
The FC800D is part of the SENCOM4.0 platform family and can connect to the FLXA402T. Thanks to a smart unit equipped behind the electrode unit, FC800D gives you more insight into the ordinary operation.
The FC800D is suitable for water distribution measurements at pumping stations, water delivery points, schools, and public facilities.
*Based on a Yokogawa survey in 2019
Continuous automatic cleaning with beads realizes contamination-resistant and stable measurement
You can reduce maintenance time dramatically thanks to easy maintenance structures.
Each process at treatment plant: from reservoir to treated water, distribution system
Water Purification Treatment
Measured Object | Free chlorine in tap water | |
---|---|---|
Measuring System | Amperometric method (Polarographic method) using rotating electrode | |
Measuring Range | Free chlorie 0-1 to 0-3 mg/L(-F1), Combined chlorine insentive version 0-1 to 0-3 mg/L(-C1), Free chlorie 0-5 mg/L (-F2) (*), Free chlorie 0-10 mg/L( -F3) (*) (*) As for "-F2" and "-F3", please refer to the General Specifications. | |
Sample Condition (-F1, -C1) | Temperature | 0 to 50°C (Temperature compensation is effective within 0 to 40°C) |
pH | No pH fluctuation pH6.5 to 7.5 (pH6.0 to 8.0 :Chlorine value 1mg/L(-F1)) | |
Conductivity | No conductivity fluctuation 100 to 300 μS/cm (100 to 500 μS/cm :Chlorine value 1mg/L(-F1)) | |
Flow Rate | 0.1- 2.5 L/min | |
Pressure | 1-150 kPa | |
SS (Suspended Solid) | under 10 mg/L | |
Power Supply | 100 to 240 V DC, 50/60 Hz | |
Digital Communication | Modbus TCP/IP, Modbus RTU | |
Installation | Indoor (Outdoor use requires protection against rain.Direct sunlight must be avoided.) |
Linearity | ±5% of reading or ±0.05mg/L whichever is greater |
---|---|
Repeatability | ±1% of reading or ±0.01mg/L whichever is greater |
* As for "-F2" and "-F3", please refer to the General Specifications.