Technical Parameters of KIFFM - EA Flow Injection Spectrophotometric Analyzer
Product Introduction
The KIFFM - EA flow injection spectrophotometric analyzer is a fully - automatic analysis and detection system specifically designed for spectrophotometric analysis. It integrates a high - precision dual - peristaltic pump numerically - controlled wide - speed - regulation flow injection sampling system, a powerful photometric analysis data acquisition system, and chemical analysis kinetic spectrogram analysis software. It can be combined with various spectrophotometers for flow injection photometric analysis.
Application Fields
Technical Features
High - precision Peristaltic Pump Wide - range Digital Speed - regulation System
High - precision peristaltic pump wide - range digital speed - regulation system: Speed - regulation range 0 - 99 levels.
Can achieve up to 10 - channel pipeline injection (5 channels/pump)
16 - channel automatic/manual valve
High - precision Data Acquisition System, Controlled by the Main Computer, Capable of
Automatic/manual zero - setting
Gain control: 1×, 2×, 4×, 8×, 16×
Filter cut - off frequency adjustment: 10 - 100 Hz
Sampling rate setting: 1 - 100T/S
External signal input interface, which can receive analog signals from other instruments, with an input range of 0 - 5V
Detector: One 722N Spectrophotometer
The 722N visible spectrophotometer adopts a single - beam optical path structure with low stray light and high resolution.
The instrument has good stability and reproducibility.
With the application of the latest microcomputer processing technology, the operation is more convenient, and the test can be completed more quickly.
It has control functions such as automatic calibration of 0%T and 100%T.
The LED digital display can show parameters such as transmittance, absorbance, and concentration, improving the reading accuracy of the instrument.
Print output, equipped with a serial port, can be directly connected to a printer to print experimental data.
Wavelength range: 325 - 1000 nm
Wavelength accuracy: ≤ ±2 nm
Wavelength repeatability: 1 nm
Spectral bandwidth: 5 nm
Stray light: < 0.1%(T) (measured with NaNO2 at 360 nm)
Stability: Dark current drift: 0.2%(T)/3min; Bright current drift: 0.5%(T)/3min
REMEX Fully Chinese - Language Data Analysis System
The measurement system can complete the setting and control of sampler parameters, the setting and control of analysis parameters, real - time data acquisition of chemical reaction kinetics, and curve plotting. It can choose to perform absorbance/intensity measurement. In the repeated measurement state, it can perform kinetic curve tests such as cumulative superposition and average superposition synchronously with the sampling system.
Measurement and processing of measured spectrogram data: The value at any point of the measured spectrogram can be read. The effective area can be set for the waveforms at any two points of the measured spectrogram, and the area, peak value, valley value, and average value can be calculated. Abnormal points in the measured spectrogram can be removed one by one or in multiple points.